Abstract
This paper introduces a new hybrid E/F power amplifier for subcutaneous applications. The proposed design is based on Amplitude Shift Keying (ASK) modulation technique with Industrial Scientific Medical (ISM) band operating at 135 kHz to avoid the tissue damage. The Hybrid E/F power amplifier combines the tuning network circuits and uses a control unit to control a selectable switch to choose the desired amplifier. The Amplitude Shift Keying modulator is proposed to supply the proposed hybrid amplifier with approximately constant DC voltage. The proposed power amplifier aims to provide an appropriate power to power the implantable devices and reduce the hardware of the subcutaneous devices depending on the specific applications
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Iraqi Journal of Computer, Communication, Control and System Engineering
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.