Abstract
A novel burning calorimeter design based on a heat pipe is presented. A circuit for automated control over operation of the proposed device is considered. The stability of the results is assessed. Several accessory parameters affecting the reproducibility of measurement results are evaluated. One of recent designs of the KTT-7 calorimeter, based on a thermosiphon, is analyzed. The uncertainty in heat power measurements on KTT-7 is evaluated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.