Abstract

An electro-magnetic interference noise shielding enclosure for Pockels cells for high speed synchronized switching has been set-up and tested. The shielding effectiveness of the aluminum enclosures housing the Pockels cells and the electronic circuitry has been measured using a high impedance probe and is found to be ∼50 dB. This ensures a noise-free and synchronized electro-optic switching in an Nd:glass re-generative amplifier of chirped pulse amplification based table top terawatt laser system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.