Abstract

We present a method for the optimal design of superconducting magnet systems for magnetic resonance imaging (MRI). The method integrates a linear-programming technique with the finite-element method (FEM) to calculate minimum-volume coil configurations subject to magnetic field homogeneity constraints for MRI systems that contain general nonaxisymmetric magnetic yoke structures. The method rapidly converges and only requires a small number of iterations and FEM analyses to be performed. In particular, the method is well suited for magnet design problems that necessitate large 3-D FEM models. We demonstrate the method with the optimal design of an open and compact 0.5 T yoked biplanar magnet assembly considered for use in an integrated medical linear accelerator and MRI system. In particular, the coil configuration for this magnet design is constructed from a MgB <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> high-temperature superconducting material that operates in a conduction-cooled cryogen-free environment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.