Abstract
The Omegatron mass spectrometer, originally developed for measuring atomic constants, was employed for measuring both positive ions and negative ions in high density, electron cyclotron resonance etching plasmas. The main disadvantages of the conventional design of the Omegatron include the observation of resonance frequency shifts and poor mass resolution in a relatively weak magnetic field. Based on 2D numerical analysis, several modifications were carried out to overcome these limitations: the (±) ion collector, differential pumping, different RF excitation methods, an additional magnet for uniform magnetic field, etc. Experimental results indicated the presence of H- and H-2 in H2 plasmas, and F- and CF- in CF4 plasmas.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.