Abstract

Current feeder system (CFS) of steady state superconducting tokamak 1 (SST-1) consists of ten pair of vapour cooled current leads (VCCL). Helium vapour consumption of a VCCL is an important parameter to be measured accurately during operation of current leads. To measure flow precisely and accurately, we use venturi type flow element. A venturi flow element was designed for required flow rate with accuracy of approximately +/-1% and the same was machined, finally it was calibrated and validated to define its discharge co efficient and found to be in agreement with the designed value. These venturi flow elements with Differential pressure transmitter (DPT) have been installed in helium gas return network of current feeder system of SST-1. This paper describes design, validation and installation and commissioning of venturi flow element with its DPT.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.