Abstract

In this work, a digital signal peripheral interface controller (dsPIC) based repetitive transcranial magnetic stimulator (rTMS) was designed and tested under low voltages. In addition, some limited knowledge of TMS, especially design parameters and notions concerned with it, also were investigated. The reason employing the dsPIC in the design is that design parameters can effectively be controlled. Pulse width modulation and switching output of the control unit, which is necessary to control the rTMS device, were controlled in a more effective way. The other novelty is that developed system can be used for therapeutic or diagnostic purposes in future work provided by digital signal processing performance of dsPIC. Bounded-cylindrical in shape head model made from nonmagnetic material, was used during the tests of the system. Spectrum analyses of clicking sounds were performed with FFT by using MATLAB. The effectiveness of the designed system have been proved by its' measurement results compared with previous works.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.