Abstract
Fourteen mild-to-moderate asthmatic patients completed a randomized four-way crossover scintigraphic study to determine the lung deposition of 200 microg budesonide inhaled from a Respimat Soft Mist Inhaler (Respimat SMI), 200 microg budesonide inhaled from a Turbuhaler dry powder inhaler (Turbuhaler DPI, used with fast and slow peak inhaled flow rates), and 250 microg beclomethasone dipropionate inhaled from a pressurized metered dose inhaler (Becloforte pMDI). Mean (range) whole lung deposition of drug from the Respimat SMI (51.6 [46-57]% of the metered dose) was significantly (p < 0.001) greater than that from the Turbuhaler DPI used with both fast and slow inhaled flow rates (28.5 [24-33]% and 17.8 [14-22]%, respectively) or from the Becloforte pMDI (8.9 [6-12]%). The deposition pattern within the lungs was more peripheral for Respimat SMI than for Turbuhaler DPI. The results of this study showed that Respimat SMI deposited corticosteroid more efficiently in the lungs than either of two widely used inhaler devices, Turbuhaler DPI or Becloforte pMDI.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.