Abstract

The present study was to evaluate the outcomes of Denali filter retrieval. We retrospectively reviewed 143 patients who received Denali filter insertion from September 2015 to April 2020. Ninety-seven patients who required removal of the filters were include in this study. Filters were retrieved with either standard or advanced techniques. Venography before and after retrieval was obtained to evaluate technical success, complications and duration of filter insertion. All 97 filters were retrieved successfully without complications. Ninety-two (94.8%) were retrieved with standard technique and 5 filters (5.2%) required the advanced technique. There were two cases with a filter angle greater than 15 degrees. Inferior vena cava penetration was shown in 17 patients (17.5%) on venography but was not associated with contrast media extravasation after filter removal. The Denali filter showed a high rate of successful retrieval without complications. This study adds value to previous studies and trials showing that the Denali filter is a reliable and safe filter that can potentially improve retrieval rates, with increasing use of this device.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.