Abstract

Quantitative myocardial PET perfusion requires decay correction (DC) of the dynamic datasets to ensure that measured activity reflects true physiology and not radiotracer decay or frame duration. DC is typically performed by the PET camera system, and the exact algorithm is buried within the settings and assumed to be correct for quantitative perfusion data. For quantitative myocardial perfusion, sequential dynamic images should be decay-corrected to the activity at the midpoint of the first scan in the sequence. However, there are different DC algorithms that can be implemented depending on the needs and expertise of the laboratory. As such, before quantitative myocardial perfusion is performed, the DC technique of a camera system should be tested.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.