Abstract

To realise gate dielectrics in the present ULSI technology ultra thin oxynitrides are used. Nitrogen quantitative depth profiles are mandatory to characterise this material. Depth resolution is the key analytical parameter to obtain useful ultra thin oxides characterisation. To improve this resolution very low primary ion impact energy is required. In the present paper, we compare quantitative depth profiles carried out by dynamic-SIMS and ToF-SIMS, respectively. Dynamic-SIMS analyses have been performed using a Cameca 4-f and the new Cameca Sc-Ultra 300 instrument. Different impact energies and incidence angles were used in combination with MCs + ion monitoring. The D-SIMS profiles at keV and sub keV primary beam impact energies, are discussed and compared with ToF-SIMS data obtained using a IONTOF IV instrument.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.