Abstract
We report the performance of capillary zone electrophoresis coupled with an electrokinetically pumped electrospray interface and an Orbitrap-Velos mass spectrometer for high sensitivity protein analysis. We first investigated the system for quantitation of the tryptic digest of BSA. The system produced outstanding linearity with respect to peak height, number of peptide IDs, and spectral counts across the range of 12 nM to 750 nM (60 amol to 3.5 fmol) of BSA injected. One peptide produced a detection limit of 0.3 nM (1.5 amol) injected. We also analyzed 700 pg of a tryptic digest prepared from a RAW264.7 cell lysate; ten proteins were identified in triplicate analyses after filtering the data with peptide confidence value as high. This sample size corresponds to the protein content of approximately ten eukaryotic cells.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.