Abstract

The wear debris caused by joint prosthesis is well known to induce an inflammation in the peripheral tissue. The authors carried out two kinds of experiments, to clarify the phagocytable size of wear particles and the cytotoxicity of macrophage related to the size and materials. The test materials were Al2O3, SiO2, TiO2 fine particles and Ti-6Al-4V, Co-28Cr-6Mo wear particles. The results showed that the phagocytable particle size was less than 11.9±11.2µm. It appears that the cytotoxicity did not depend on the particle size, even if the particles were phagocytable size. In the relationship between material type and inflammation, damage levels were found to be different between SiO2 and TiO2 particles, even if neither material released metal ions. The cells were damaged more severely by SiO2 particles than by Co-28Cr-6Mo for which the eluted ion could not be ignored (damage levels; SiO2>Co-28Cr-6Mo>Ti-6Al-4V>Al2O3≧TiO2). For these reasons, it was confirmed that there was a factor in addition to the toxicity of the eluted metal ion (i. e., the implant material's corrosion resistance ability), which influenced the inflammation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.