Abstract

The purpose of this study was to evaluate the cyclic fatigue resistances of the WaveOne Gold (WOG), ProTaper Next (PTN), 2Shape (TS) instruments. Totally 45 new WOG (25/.07), PTN (25/.06), TS (25/.06) files were tested inside the artificial canal of a custom-made stainless steel block with an inner diameter of 1.5 mm, a 60° angle, and a 5-mm radius of curvature. A 16-mm-long file segment (from the tip) was introduced and was immersed in 37°C water. The number of cycles to fracture (NCF), time to failure (TTF), fractured fragment length (FL) was recorded and the fractured surface was examined using microscope. WOG > PTN > TS according to TTF results (P = 0.00). PTN > TS according to NCF results (P = 0.00). The FL values showed no significant difference (P = 0.335). Reciprocating motion can be used more safely than continuous rotation due to the higher cyclic fatigue resistance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.