Abstract

Background Adequate compliance with the existing guidelines for cleaning and disinfection of gastrointestinal endoscopes and accessories is necessary to obtain high-level disinfection and prevent pathogen transmission. Aim To investigate cleaning and disinfection practice in China. Methods A questionnaire with 21 questions concerning gastrointestinal endoscopy reprocessing was sent by e-mail to 189 endoscopy units in China. Results One hundred and twenty-two (80.39%) of the 189 units responded. Compared with the low-workload units (<5000 procedures/year), the high-workload units (≥5000 procedures/year) had a significantly higher number of gastrointestinal endoscopes (25.8 ± 3.6 vs. 4.7 ± 1.8, p < 0.01) and the higher possessing rate of automated endoscope reprocessors (43.9% vs. 3.1%, p < 0.01). Glutaraldehyde was the most commonly employed disinfectant (88.5%) in all the units. In 23/122 (18.8%) units, the exposure time to glutaraldehyde was <45 min in the case of infectious disease patients. Eighty-six of 122 (70.5%) units reused disposable materials, of which 21/86 (24.4%) reused disposable forceps and disposable polypectomy hooks, and 2/86 (1.6%) reused disposable injection needles intermittently. Conclusion Although gastrointestinal endoscopy has developed rapidly in China in the past decade, there is still room for improvement in the practice of endoscopy reprocessing, especially in middle-sized and small cities.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.