Abstract
The purpose of this paper is to review the current role of computed tomography (CT) in the imaging of left ventricular assist devices (LVAD). Recent studies have highlighted the prognostic role of measuring skeletal muscle CT attenuation in predicting adverse LVAD outcomes. CT is also useful for evaluation of inflow cannula malposition, and outflow graft complications such as obstruction and kinking. In addition, 18F-fluorodeoxyglucose positron emission tomography-computed tomography (FDG PET-CT) has been shown to be of value in evaluation of suspected LVAD infection. Newer CT techniques such as dual-energy CT and metal artifact reduction algorithms can potentially generate better quality CT images, though currently there is no high-quality published literature on their use in LVAD imaging. CT is a useful technique as part of a multi-modality approach in the evaluation of LVADs and associated complications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.