Abstract

The purpose of this study was to evaluate the effects of target diameter and display-field of view (D-FOV) in modulation transfer function (MTF) by circular edge strategy using the computed tomography (CT) image measurement program "CTmeasure". We calculated the MTF (MTF(edge)) using the circular edge strategy applied to cylindrical phantom (200 mmφ) that inserted with cylinders have 10, 20, 30, and 40 mm diameters. The phantom images were reconstructed using filtered back projection method varied with D-FOV (240, 320, 400, and 500 mm). The study compared both MTF(edge) and MTF(wire) at MTF50% and MTF(10%) for target diameter and D-FOV, respectively. The MTF(edge) by the different of target diameter indicated in rough compatibility. However, MTF(edge) of D-FOV diameters (320, 400, and 500 mm) decreased in the high frequency range. The circular edge strategy for MTF depended on the D-FOV, however, it was little dependent on target diameter using the CT image measurement program "CTmeasure".

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.