Abstract

A new HPTLC method was envisaged to determine rosmarinic acid (RA) in different matrices with the aim of testing the influence of optimizing the main HPTLC operative parameters in view of a more stringent validation process. HPTLC LiChrospher silica gel 60 F254s, 20 cm × 10 cm, plates with toluene:ethyl formate:formic acid (6:4:1, v/v) as the mobile phase were used. Densitometric determinations were performed in reflectance mode at 330 nm. The method was validated giving rise to a dependable and high throughput procedure well suited to routine applications. RA was quantified in the range of 132-660 ng with RSD of repeatability and intermediate precision not exceeding 2.0% and accuracy within the acceptance limits. The method was tested on several commercial preparations containing RA in different amounts.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.