Abstract

When testing the Microgenics CEDIA assay for immunological buprenorphine analysis, cross-reactivity between the buprenorphine reagents and opiates was observed at concentrations higher than 120 mg/l morphine, 320 mg/l methadone, 30 mg/l codeine, 60 mg/l dihydrocodeine and 520 mg/l morphine-3-glucuronide. The cross-reactivity with morphine has the greatest impact on routine screening as opiate maintenance therapy in Austria is also performed with slow-release oral morphine. The use of a second cutoff value of 30 mug/l for urine samples that are (immunologically) positive for opiates is therefore suggested, compared to the cutoff value of 5 microg/l proposed by the manufacturer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.