Abstract
Several developments were made in the GATE simulation platform to allow accurate modeling of the count rate performances of PET scanners over a wide range of activity concentrations. A background noise module, a dead time and limited bandwidth modeling for the coincidences, and a delayed coincidence builder were added in the code. The results obtained for the modeling of the ECAT HRRT and Focus 220 scanners with the newly developed modules are discussed. They show that GATE can be used to accurately simulate the single event, prompt coincidence and delayed coincidence rates, from very low activity levels in the field of view up to levels that saturate the acquisition system. The new developments were committed into the public release of GATE, making them available for the whole community, thanks to the open source license under which GATE is published (LGPL).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.