Abstract
We report flip angle and spoiled gradient echo measurements at 7 Tesla on human brains in three-dimensional imaging, using strongly modulating pulses to counteract the transmitted radiofrequency inhomogeneity problem. Compared with the standard square pulse results, three points of improvement are demonstrated, namely: (i) the removal of the bright center (typical at high fields when using a quadrature head coil), (ii) the substantial gain of signal in the regions of low B(1) intensity, and (iii) an increased 35% signal uniformity over the whole brain at the flip angle where maximum contrast between white and gray matter occurs. We also find by means of simulations that standard BIR-4 adiabatic pulses need several times more energy to reach a similar performance at the same field strength.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.