Corrosion of the head-neck junction of implants used in total hip arthroplasty is a complex problem. Clinical severity appears to be multifactorial, and the predictive variables have yet to be consistently identified in the literature. Corrosion should be considered in the differential diagnosis of hip pain following total hip arthroplasty regardless of the type of bearing surface used. The most common presentation, pain followed by instability, is similar to complications associated with metal-on-metal articulations. The diagnosis of implant corrosion of the head-neck junction can be challenging; an infection workup should be performed along with analysis of serum metal ion levels and cross-sectional imaging. In the short term, a well-fixed stem may be retained, and the exchange of an isolated head with a ceramic femoral head seems to be a promising option for certain implants. Further research with longer follow-up is warranted, and high levels of evidence are needed to determine whether this approach is generalizable.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.