Abstract
A sterling silver tracheostomy tube broke in use resulting in the death of a child. The nature of the fracture in the brazed joint is described, and corrosion identified as the mechanism of degradation of the joint. The corrosion is due either to exposure to body fluids over a long period, or to the use of hypochlorite solutions for cleaning and disinfection. The silver braze used had inferior corrosion resistance. It is recommended that appropriate grades of silver brazing alloy be used in the future, and that the mechanical integrity of medical devices joined with silver braze be regularly checked to anticipate failure in use.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.