Abstract
The assessment of coagulating blood elasticity has gained importance as a result of several studies that have correlated it to cardiovascular pathologic conditions. In this study we use supersonic shear wave imaging (SSI) to measure viscoelastic properties of blood clots. At the same time, classical rheometry experiments were carried out on the same blood samples taken within the first few seconds of coagulation. Using SSI, phase velocities of the shear wave indicated increasing dispersion with time. In all cases, the frequency bandwidth of propagating shear waves changed from 20-50Hz at the first few min of coagulation to around 300Hz toward the end of experiments. Using the values of G' and G″ from the rheometry studies, the theoretical shear wave velocities were calculated and correlated with SSI measurements. Results of the two techniques were in very good agreement, confirming that SSI provides accurate measurements of viscoelastic properties as corroborated by conventional rheometric measurements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.