Abstract

This paper demonstrates, in summary form, some corrections to be made to values of spectral flux as measured with Eppley spectral pyranometers. The results of an investigation into the occurrence of what are believed to be systematic errors in scaled hemispherical thermopile systems, employed in pyranometer design, are presented. This investigation was prompted by an analysis of a large mass of solar spectral measurement material, which indicated an apparent change in spectral pyranometer sensitivity under conditions of continuous operation. The study also included pyranometers of natural- and forced-ventilated design. The results indicate that the corrections so far established ought to be of general application, within practical limits, to the present Eppley spectral pyranometer and, perhaps with some modifications, to other constructions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.