Abstract

A 1974-vintage 25 cm field of view gamma camera was converted to a bone mineral densitometer using a purpose-built converging collimator with attached C-arm holding a 250 mCi gadolinium (Gd) 153 source. After addition of a second pulse-height analyser, increasing the high voltage gain and connecting to a digital computer, spinal and hip bone mineral density measurements were made. Accuracy of 2% and long-term reproducibility of 2.7% were obtained using an anthropomorphic bone phantom. For lumbar spine measurements in normal volunteers, long-term reproducibility was 3.2% and for osteoporotic (spinal fracture) patients (mean density 831 mg cm-2) reproducibility was 3.7%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.