Abstract

An integration of SENSitivity Encoding (SENSE) with continuously moving table (CMT) MRI for extended field-of-view (FOV) acquisitions is described. In this work, the approach in which receiver coils are attached to the object and move in synchrony with the scanner table is considered. Technical issues dealing with the implementation of SENSE-CMT are addressed, including coil calibration, correction for non-uniform magnetic gradients, and specific reconstruction steps. An explanation of combining SENSE with gradient non-linearity correction is given, as the latter becomes necessary in CMT acquisitions where a large sampling FOV is used. It is hypothesized that SENSE can provide at least a 2-fold improvement in lateral spatial resolution compared to non-accelerated CMT acquisitions. The hypothesis is tested in phantoms, where the effectiveness of both SENSE and gradient non-linearity correction to improve spatial resolution is shown. The SENSE-CMT technique is further demonstrated in vivo with contrast-enhanced MR angiography of the peripheral vasculature.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.