Abstract

We have used a glucose oxidase-based sensor implanted in the striatum of freely moving rats to determine the concentration of extracellular glucose in two distinct ways. With a modification of the zero net flux method, in which different concentrations of glucose are infused through a dialysis probe glued to the biosensor, we calculated the concentration at which there was no change in glucose current by regression analysis; this gave a concentration of 0.351 +/- 0.016 mM. Calculating the concentration from the basal current and the in vitro calibration of the biosensor was not significantly different from this. The basal extracellular glucose concentration determined by either method remained constant over a period of several days. Infusion of 50 microM veratridine through the adjacent dialysis probe caused a steep decrease in glucose current as soon as the drug reached the brain in contrast to the delayed fall (7.5 min) seen with microdialysis in previous experiments from this laboratory. These results demonstrate that this biosensor provides a direct, real-time measure of the extracellular concentration of glucose.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.