Abstract
Conventional intraoperative nerve monitoring, predicated on intermittent stimulation, can predict recurrent laryngeal nerve (RLN) palsy only after the damage has been done. Fifty-two patients (52 nerves at risk) who underwent continuous intraoperative nerve monitoring (CIONM) for thyroid surgery via vagus nerve stimulation had their electromyographic (EMG) tracings recorded and correlated with surgical maneuvers and postoperative RLN function. There was 1 imminent loss of signal (LOS) with intraoperative signal recovery and there were 4 losses of signal with corresponding unilateral transient RLN palsy. When EMG amplitude decreased >50% and EMG latency increased >10%, LOS and postoperative RLN palsy were noted in 4 of 8 patients (50%) who had multiple combined events. In 9 of 13 patients (70%) who developed adverse EMG changes, modification of the causative surgical maneuver resulted in recovery of those EMG changes and aversion of impending RLN palsy. CIONM reliably signaled impending nerve injury, enabling immediate corrective action.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.