Abstract
Abstract We describe a system that monitors head movement continuously and corrects the data to remove the effects of motion. Head position is monitored by real-time analysis of signals of three small coils mounted on the subject's head. Data correction is done off-line. Tests of the system with both dipole fitting and SAM beamformer analysis show that it allows correction of errors in source locations due to head movement to ∼ 1 mm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.