Abstract

In a typical continuous-flow optical pumping setup, the chemical shift of xenon in the adsorbed phase depends on the gas flow rate due to warming of the sample surface by the gas stream. Calibration of the system using the (207)Pb resonance of solid lead nitrate is necessary to determine the actual sample temperature. Optimum pulse repetition rates are strongly affected by gas flow and spin-lattice relaxation rates. The interplay of flow and pulse repetition rate alters signal intensity ratios and may lead to the complete suppression of signals.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.