Abstract
Water sterilization at sterility assurance levels exceeding 10−6 has been achieved with a novel high temperature and pressure sterilization system (HAPSS) capable of continuous-flow sterilization. The tested sterilizer produces >2,000 L/day and does not require filters, moving parts, or regular maintenance and can operate for multiple years with only water and standard voltage/amperage electricity. Sterility assurance levels were determined using Geobacillus stearothermophilus spores with inactivation experiments as a function of temperature and flow rate (i.e., duration/dosage at given temperature). Sterility of 1.0 × 104 CFU/mL G. stearothermophilus spore suspensions was achieved at the highest water flow rate of 1.4 L/min at 140 and 130 °C. The low cost and maintenance-free operation of HAPSS is envisioned to impact water sterilization needs of developing nations, hospitals, and commercial point of use applications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.