Abstract
A new prototype of contactless conductivity detector, smaller and easier to operate than the former version, is described. For a fused-silica capillary with 142-μm wall thickness and voltages up to 25 kV, it can be placed at the low- or high-voltage end of the column. This feature allowed implementation of an apparatus with sample introduction at the grounded end of the column. The input signal is an important parameter for determining the signal-to-noise ratio ( S/ N) of the detection system. An optimization procedure of its amplitude and frequency is proposed. Although the S/ N must be determined by introduction of actual samples, the operating conditions can be optimized merely by changing the signal parameters and by using a mathematical procedure. Thus, an easy and fast optimization routine can be carried out. Mathematical and instrumental backgrounds are discussed, and experimental support of the technique’s effectiveness is presented.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.