Abstract
Conventional Johnson noise thermometers based on switching correlators have conflicting matching requirements for the sensing resistance. To mitigate distortion effects in the correlator, the RT products of the two sensors must be the same, and to mitigate frequency-response errors in nominally identical input circuits, the two sensing resistances should be the same. A noise thermometer using synthetic noise for the primary reference signal overcomes this conflict because the output voltage and output resistance are independent. This paper presents the rationale and design constraints for a noise thermometer using a synthetic-noise source based on Josephson junctions. The quantized voltage noise source developed at NIST produces tunable waveforms with a spectral density composed of a comb of frequencies of equal amplitude and random phase. In addition to the conventional noise-power and impedance constraints, it has additional constraints relating to the number of tones and the tone spacing.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.