Abstract
A long-term comparison was done between the manufacturer-stated 192Ir activity and the measured 192Ir activities determined with a well-type ionization chamber. Sources for a Nucletron Micro Selectron high-dose-rate (HDR) unit were used for this purpose. The radioactive sources reference activities were determined using a PTW well-type ionization chamber traceable to the National Institute of Standards and Technology Primary Calibration Laboratory. The measurements were taken in a period of 56 months with 17 different radioactive sources. The manufacturer stated activities were taken from the source calibration certificate provided by the manufacturer. These values were compared with the measured activities. The results have shown that both the percentage deviation of the monthly control measurements with the well-type chamber and the ratio between the measured activities to the manufacturer-stated value lie within ± 2.5%. These results were compared with similar published data and with uncertainty level (3% of the mean and 5% maximum deviation from mean) for brachytherapy sources calibration recommended by the AAPM. It was concluded that a threshold level of ±2.5% can be used as a suitable quality assurance indicator to spot problems in our department. The typical ±5% uncertainty as provided by the manufacturers may be tightened to ±3% to be more in line with published AAPM reports.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.