Abstract

Different quantitative assays for HCV-RNA are available that report test results in International Units (IU)/ml based on the WHO International Standard. Thus, assays have been calibrated with standard material containing HCV genotype 1, so the consistency of quantitation might differ between assays for other HCV genotypes. Three commercial HCV nucleic acid amplification techniques (HCV-NAT) were compared for quantitation consistency of genotype 4 and 1 specimens from an Egyptian blood donor panel (n = 92). Consistency of quantitation between HCV-NATs was higher for genotype 1 than for genotype 4. Most quantitative results reported by the assays were in the same range, but some genotype 4 samples were missed by two of the assays. The Abbott assay showed higher concentrations for genotype 4 than the two Roche assays. Follow-up investigations of individuals should use the same assay unless another assay has been validated properly. Standardization of HCV-NAT assays remains an issue.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.