Abstract

To analyze the consistency of effective orifice area (EOA) of prosthetic mitral valve estimated using 2- dimensional (2D) and 3-dimensional (3D) transesophageal echocardiography (TEE). This study was conducted among 34 patients undergoing mitral valve replacement surgery in Nanjing First Hospital between March and June in 2019. The diameter of the left ventricular outflow tract (LVOT) measured by 2D-TEE was used to calculate the cross sectional area of LVOT (CSALVOT). In 3D-TEE method, LVOT area was measured directly by planimetry on an enface view. The EOAs of the prosthetic mitral valve were calculated for both methods using the continuity equation. Bland-Altman plot consistency test was used to analyze the consistency between the two sets of EOA results, and linear regression analysis was used to analyze their correlation. The EOA of the prosthetic mitral valve differed significantly between 2D method and 3D method (2.22±0.71 cm2 vs 2.35±0.70 cm2, P < 0.001) with a mean difference of -0.14±0.20 cm2 and 95% coherence boundaries of (-0.53, 0.25 cm2). The regression equation for EOA-3D and EOA-2D is y=0.27 + 0.94x, showing a good correlation between the two methods. EOA estimation of the prosthetic mitral valve using 2D and 3D TEE has a good consistency, and the results estimated by the 2D method are slightly lower by about 6% than those by the 3D method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.