Abstract

By regarding the dialysate flow in a hollow fiber bundle as a flow in a porous medium, the average dialysate pressure of several hollow fibers is numerically determined using the transverse conductivity perpendicular to the hollow fibers and longitudinal conductivity parallel to the hollow fibers. Determination of the dialysate pressure isobars becomes possible if the transverse-to-longitudinal hydraulic conductivity ratio can be obtained experimentally. Since the dialysate fluid flows in the transverse direction at the inlet and outlet, the radial dialysate pressure gradient is enhanced at the inlet and outlet if the transverse conductivity is low. However, even if the transverse conductivity is low, the use of a slender dialyzer suppresses the radial dialysate pressure gradient. If the transverse-to-longitudinal hydraulic conductivity ratio is lower than one-tenth the radius-to-total-length ratio of the cylindrical dialyzer, the radial dialysate pressure gradient is negligible compared to the central axial dialysate pressure gradient.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.