Abstract

A computer-assisted analysis of coronary obstructions from cineangiograms is presented, which includes both the geometric and hemodynamic evaluation of coronary stenosis severity. Single frame images are digitized into a 512×512×8 bit array after a cine-to-video conversion. Automatic edge tracking is performed using a combination of derivative and threshold methods. Vessel borders from two orthogonal views of the arterial segment are used to create a three-dimensional reconstruction of the stenosis, which serves as a basis for calculation of absolute and relative geometric dimensions, stenotic resistance and trans-stenotic pressure gradient for various given flow rates. A preliminary performance evaluation of this method was made by analysis of x-ray phantoms representing stenoses of known dimensions, which were filmed under quasi-clinical conditions. The results are discussed with respect to their accuracy and reproducibility.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.