Abstract
Intramedullary fixation for distal radius fractures is reported to be free of hardware irritation and less invasive than other fixation methods. Some specific complications associated with intramedullary fixation, such as radial nerve sensory neuritis, have been reported, but no study has focused on the complication rates of intramedullary fixation for distal radius fractures in the elderly population. Furthermore, no studies have analyzed common complications, such as carpal tunnel syndrome and flexor tenosynovitis including trigger finger, among patients with distal radius fractures treated by intramedullary fixation based on a comprehensive complication checklist. We reviewed the medical records of 52 elderly patients with distal radius fractures treated with intramedullary nail fixation. We investigated the postoperative complications in these patients using McKay's complication checklist. 5 patients experienced radial nerve sensory disorder, and one patients developed carpal tunnel syndrome. All neurological symptoms resolved spontaneously, and these neurological complications were categorized as mild. Further, 3 patients developed trigger finger at the A1 pulley and needed triamcinolone injections for symptomatic relief. There were no tendinous complications around the implanted hardware. All tendinous complications were categorized as moderate complications and resolved with steroid injection therapy. Among skeletal complications, 1 case of postoperative volar displacement resolved with good functional outcome without the need for corrective osteotomy. This was considered a mild complication. The total complication rate was 19.2%. All complications were categorized as mild or moderate, and no patients experienced severe complications that needed further surgery such as hardware removal. Intramedullary fixation for distal radius fractures was free from tendinous complications such as tenosynovitis and tendon ruptures around the implant, which are frequently caused by volar locking plate fixation. However, this less invasive technique could not avoid common complications such as trigger finger and carpal tunnel syndrome associated with distal radius fractures.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.