Abstract

This paper presents a new method for the determination of Hg(II) in waters, using “indirect” spectrophotometric measurements of the Hg(II) dithizone complex. The reagent is a mixture of dithizone, CDTA (1,2-cyclohexylene-dinitrilo-tetraacetic acid), thiourea, ethylic alcohol and glycine. Sensitivity and working range are 0.4 μg · l−1 and 2–100 μg · l−1, respectively. A comparison of the results with those of the cold-vapor atomic absorption spectrometry shows a good correlation for different samples revealing only small interferences. The new method is well suited for on-line measurements, easy to run and has low cost equipment.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.