Abstract

An anti-Jka was not detectable by the DiaMed-ID Micro Typing System but was strongly reactive when tested using other indirect antiglobulin test (IAT) methods. However, it was detectable in the DiaMed-ID system when the recommended diluent ID-Diluent 2, was replaced by other low ionic Strength saline (LISS) solutions. This led us to believe that the problem lay with the ID-Diluent 2. The anti-Jka was investigated using various low ionic strength diluents in the IAT against both polyspecific and monospecific antiglobulin reagents. We tested the ID-Diluent 2 for its ability to inhibit complement activation, and bind calcium. The anti-Jka was shown to be a 'complement-only' binding antibody that reacted in all IATs performed except when tested with ID-Diluent 2. The ID-Diluent was shown to prevent haemolysis of ABO lytic antibodies, and further testing shows that it binds calcium. The use of a diluent which prevents haemolysis and binds calcium in an IAT system that contains anti-complement as well as anti-IgG must be questioned.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.