Abstract

The humidity output of heated humidifiers may be compromised by inlet gas temperatures exceeding approximately 26 degrees C, with humidity dropping below the recommended levels for intubated patients. A new version of the Fisher & Paykel MR850 humidifier claims to deal with this problem by offering a humidity compensation option. The present study tested this feature by measuring humidity output using the gravimetric method and a hygrometer at different inlet gas temperatures (16.6 degrees C to 40.0 degrees C) with compensation on and off. It was found that the compensation is effective in maintaining humidity levels despite high inlet gas temperatures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.