Abstract
The humidity output of heated humidifiers may be compromised by inlet gas temperatures exceeding approximately 26 degrees C, with humidity dropping below the recommended levels for intubated patients. A new version of the Fisher & Paykel MR850 humidifier claims to deal with this problem by offering a humidity compensation option. The present study tested this feature by measuring humidity output using the gravimetric method and a hygrometer at different inlet gas temperatures (16.6 degrees C to 40.0 degrees C) with compensation on and off. It was found that the compensation is effective in maintaining humidity levels despite high inlet gas temperatures.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.