Abstract

To evaluate whether video head impulse test (vHIT) gains are dependent on the measuring device and method of analysis. Prospective study. vHIT was performed in 25 healthy subjects using two devices simultaneously. vHIT gains were compared between these instruments and using five different methods of comparing position and velocity gains during head movement intervals. The two devices produced different vHIT gain results with the same method of analysis. There were also significant differences in the vHIT gains measured using different analytical methods. The gain analytic method that compares the areas under the velocity curve (AUC) of the head and eye movements during head movements showed lower vHIT gains than a method that compared the peak velocities of the head and eye movements. The former method produced the vHIT gain with the smallest standard deviation among the five procedures tested in this study. vHIT gains differ in normal subjects depending on the device and method of analysis used, suggesting that it is advisable for each device to have its own normal values. Gain calculations that compare the AUC of the head and eye movements during the head movements show the smallest variance.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.