Abstract

Rapid removal of small molecules during hemodialysis places an acutely ill patient with kidney failure at an increased risk of hemodynamic instability and for dialysis disequilibrium syndrome. The use of high-flux, high-efficiency (HEF) dialyzers may increase this risk despite reductions in blood and dialysate flow. We performed in vitro experiments to compare urea clearance at low dialysate flow and various blood flows using a low-efficiency low-flux (LEF) and a HEF membrane. Compared to LEF, there was a significant increase in the clearance of urea at all blood flows with the HEF (all P values < 0.005). HEF dialyzer (F180NR) had higher urea clearance at a blood flow of 150 mL/min than LEF dialyzer (F5) at blood flow of 300 mL/min (144.1 ± 0.99 vs. 130.1 ± 0.001 mL/min for F180 vs. F5, respectively, P < 0.002). Our data suggest that use of HEF dialyzer are not as safe as LEF in high-risk acute dialysis patients since these are associated with more rapid removal of urea despite reduction in blood and dialysate flow as compared to LEF.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.