Abstract

Antimicrobial susceptibility testing of cystic fibrosis (CF) isolates of Pseudomonas aeruginosa is difficult because the organisms are often mucoid and slow-growing. This study of 498 CF strains examined the correlation of results derived from two commonly used commercial systems (Vitek, MicroScan-WalkAway) with a reference method for 10 antimicrobials. Correlation to reference results was unacceptably low for all agents and both commercial systems had a high rate of very major (false-susceptible) errors. Although mucoid strains produced a 4.8% greater intermethod error, it was not markedly different than non-mucoid strains for the Vitek System. Overall, these tested commercial systems performed poorly for CF isolates in contrast to earlier reported, high correlations with the reference methods (broth microdilution frozen panels and agar dilution) of the National Committee for Clinical Laboratory Standards, the standardized disk diffusion test, and the Etest (AB BIODISK, Solna, Sweden).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.