Abstract
This study compared the damage scores and damage patterns in 19 tibial inserts from rotating hinge (RH) implants with 19 inserts from highly constrained (HC) implants. Each insert was divided into 16 damage zones and each zone was subjectively graded from a scale of 0–3 for seven different damage modes. The overall damage scores were comparable for the two groups (RH: 64.1 ± 15.4; HC: 66.1 ± 29.0; P = 0.59). The HC group, however, had greater post damage (compared to the post-hole of RH) while the RH group had greater backside damage. The pattern of damage was also different, with burnishing and cold flow being more common in HC group while pitting, scratching and embedded debris were more common in the RH group.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.