Abstract
To compare visual field test results of Glaufield Lite AP901 CTS 133 (Appasamy Associates, Mannadipet Commune, Thirubhuvanai, Puducherry, India, hereafter Glaufield Lite) with Humphrey Field Analyser (HFA, Carl Zeiss Meditec, Dublin, California, USA, hereafter HFA). A pilot study at a tertiary eye centre involving 23 normal and 24 glaucoma patients who underwent two consecutive visual field tests on (i) HFA 24-2 SITA Fast and (ii) Glaufield Lite Quick Central program. The mean testing time on HFA was significantly shorter than Glaufield Lite (normals: HFA: 2.75 ± 0.49min, Glaufield Lite: 6.85 ± 0.86min, p < 0.001; glaucoma patients: HFA: 3.45 ± 1.08min, Glaufield Lite: 6.95 ± 0.54min, p < 0.001). Reliability criteria were similar, but false-positivity was lower with Glaufield Lite. Bland-Altman analysis showed poor agreement for mean deviation (MD), [~ 2.69 units less for HFA], and acceptable agreement for pattern standard deviation (PSD) [~ 0.426 units more for HFA] between the two devices. Both perimetric techniques showed reliable test results though test duration was longer with Glaufield Lite perimetry. The MD showed poor agreement, likely due to different scales and principles used for perimetry. The PSD showed acceptable agreement, making it valid for use in glaucoma, though a direct comparison of fields from the two devices is not possible. We recommend using the same perimetry device for follow-up evaluation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.