Abstract
A comparison of 4 amino phases representing 3-aminopropyl silica batches with different amount of grafted functional groups was made. To evaluate the chromatographic properties of the adsorbents, the Tanaka test for hydrophilic stationary phases was used, and the retention of polar substances of various classes in HILIC mode was also studied. It is shown that even small changes in nitrogen content between different batches of 3-aminopropyl silica have a significant effect on the retention of polar analytes. Hydrophilicity of a substrate was shown to have the greatest effect on its chromatographic behavior, and its evaluation using the Tanaka test is the basis for a batch selection either for separation of particular classes of polar substances or for further modification aiming at obtaining new phases.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.