Abstract

To measure the lower tear meniscus height (LTMH) with image by Tearscope and using slit-lamp biomicroscopy with or without fluorescein staining. The lower tear meniscus height was measured in 31 eyes of 31 healthy volunteers (mean age 31.29 years; SD 4.83) with Tearscope and slit lamp with and without fluorescein staining. The mean values of LTMH were measured with different methods and the repeatability of the measurements was statistically analyzed. Measurement of fluorescein tear film break up time (FBUT) and Schirmer test were also performed. The mean value of LTMH was 0.21 mm (SD 0.07) with Tearscope, 0.20 mm (SD 0.06) with slit lamp without fluorescein staining, and 0.23 mm (SD 0.07) with fluorescein staining. There was no significant difference among the mean values of LTMH with these 3 different measurement methods. Based on the random effect analysis of variance (p=0.05), the measurements were more repeatable with Tearscope-plus than with the other 2 methods. No correlation was found between the LTMH and the results of Schirmer test or FBUT. The LTMH could be reliably measured with slit-lamp biomicroscopy without staining, but the measurements were more repeatable with Tearscope.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.